FDA Adverse Event
Malfunction
Summary report: N
VIDAS
MDR report key: 331768
·
Received May 8, 2001
Report
- Report Number
- 1950204-2001-00004
- Event Type
- Malfunction
- Date Received
- May 8, 2001
- Date of Event
- March 22, 2001
- Report Date
- May 4, 2001
- Manufacturer
- BIOMERIEUX, INC.
- Product Code
- JZT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
A FIELD SERVICE ENGINEER WAS DISPATCHED TO THE SITE AS PART OF A WORLDWIDE CHECK FOR CLOGGED PUMPS ON VIDAS INSTRUMENTS. DURING ONE SUCH VISIT TO THE CUSTOMER BELOW BY A BIOMERIEUX FIELD SERVICE ENGINEER THE INSTRUMENT FAILED PERFORMANCE TESTING WHICH INDICATED THAT THE PUMP IN THE FAILING POSITION COULD BE CLOGGED. A CLOGGED PUMP CAN RESULT IN INACCURATE TEST RESULTS BEING REPORTED BY THE LABORATORY. THE PUMP WAS REPLACED BY THE FIELD SERVICE ENGINEER.