FDA Adverse Event Malfunction Summary report: N

9F PRUITT OUTLYING F3 SHUNT WITH T-PORT

MDR report key: 3317245 · Received August 6, 2013

Report

Report Number
1220948-2013-00014
Event Type
Malfunction
Date Received
August 6, 2013
Date of Event
July 1, 2013
Report Date
July 22, 2013
Manufacturer
LEMAITRE VASCULAR
Product Code
MJN
PMA / PMN Number
K051067
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WE HAVE RECEIVED THE DEVICE FOR THE EVALUATION, BUT WE WERE NOT ABLE TO CONFIRM THE FAILURE MODE. ALL BALLOONS WERE FUNCTIONAL AND PERFORMED AS EXPECTED. NO PROBLEMS FOUND. LOT HISTORY RECORDS REVIEW DID NOT REVEAL ANY DISCREPANCIES RELATED TO THE COMPLAINT EVENT EITHER DURING THE MANUFACTURING OR PACKAGING PROCESSES. ALL MANUFACTURING AND QC STEPS WERE COMPLETED IN ACCORDANCE TO MANUFACTURING INSTRUCTIONS. THEREFORE, THE ROOT CAUSE(S) OF THE FAILURE REMAINS INCONCLUSIVE. HOWEVER, LEMAITRE VASCULAR, INC. HAS INITIATED CORRECTIVE AND PREVENTIVE ACTION (PLEASE REFERENCE LEMAITRE CAPA (B)(4)) TO IMPROVE BALLOON CONCENTRICITY AND DEFLATION TIME. PLEASE NOTE THAT NO PT INJURY HAPPENED DURING THIS INCIDENT.

Description of Event or Problem · 1

THE COMMON BALLOON WOULD NOT DEFLATE AT THE END OF THE PROCEDURE. THE PHYSICIAN WAS ABLE TO SAFELY REMOVE BALLOON FROM THE PT. NO PT INJURY HAPPENED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
369193 9F PRUITT OUTLYING F3 SHUNT WITH T-PORT PRUITT F3 SHUNT MJN LEMAITRE VASCULAR 2012-10 PFT2472

Patients

Seq Age Sex Outcome Treatment
1 Other