FDA Adverse Event
Summary report: N
SURGICAL TABLE
MDR report key: 33168
·
Received May 23, 1996
Report
- Report Number
- MW1009151
- Date Received
- May 23, 1996
- Date of Event
- May 8, 1996
- Report Date
- May 22, 1996
- Manufacturer
- SKYTRON DIV. OF THE KMW GROUP, INC.
- Product Code
- BWN
- Adverse Event
- Yes
- Report Source
- Voluntary report
- Reporter Location
- VA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
SURGICAL BED TILTED OVER WHEN POSITIONED FOR LITHOTOMY. WENT THROUGH THE DISTRIBUTOR TO CONTACT CO. STAFF HAS NO WAY OF KNOWING LEGS (BRAKES) HAVE LIFTED OFF GROUND. PROBLEM IS AVERTED IN ONE OF TWO WAYS. USE THE SMALLER TRANSPORT LEG SECTION OR PUT SPACERS IN HYDRAULICS TO PREVENT TABLE FROM GOING SO LOW. THE CO SAYS IT WILL SEND OUT LETTERS TO DIRECT USERS TO USE TRANSPORT BOARDS. THIS FACILITY HAS CHOOSEN TO HAVE THE SPACERS INSTALLED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SURGICAL TABLE | SURGICAL TABLE | BWN | SKYTRON DIV. OF THE KMW GROUP, INC. | 6500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO |