FDA Adverse Event Summary report: N

CO2 MONITOR

MDR report key: 33091 · Received May 13, 1996

Report

Report Number
33091
Date Received
May 13, 1996
Date of Event
April 23, 1956
Report Date
May 13, 1996
Manufacturer
NELLCOR PURITAN BENNETT, INC.
Product Code
CCK
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
WA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING OP SURGICAL PROCEDURE, MULTIPARAMETER MONITORING MACHINE FROZE UP. WAS INCAPABLE OF MONITORING THE PT DURING SURGERY. BIOMED CALLED STAT FOR MONITOR REPLACEMENT. NO INJURY TO PT. MONITOR REMOVED FROM INVENTORY AND MFR CONTACTED. BIOMEDICAL ENGINEERING STATEMENT SUGGESTS ASSOCIATION WITH MONITOR FAILURE TO THE FIRING OF ELECTRO SURGICAL UNIT (ESU) WHICH WAS USED IN SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CO2 MONITOR CO2 MONITOR CCK NELLCOR PURITAN BENNETT, INC. N-2500 NA

Patients

Seq Age Sex Outcome Treatment
1 NO INFO