FDA Adverse Event Malfunction Summary report: N

COBAS INTEGRA 400 PLUS

MDR report key: 3301221 · Received August 21, 2013

Report

Report Number
1823260-2013-05072
Event Type
Malfunction
Date Received
August 21, 2013
Date of Event
August 8, 2013
Report Date
September 17, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K951595
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.

Additional Manufacturer Narrative · 1

THE CUSTOMER INDICATED THE INSTRUMENT IS RUNNING "GREAT" AFTER THE SERVICE VISIT.

Additional Manufacturer Narrative · 1

THE INVESTIGATION COULD NOT DETERMINE A SPECIFIC ROOT CAUSE. IT WAS NOTED THERE MAY HAVE BEEN A LOOSE CONNECTION (PLUG ABS LAMP TO PCB ABS MODULE) WHICH DAMAGED THE PARTS. WHEN AN ELECTRONIC PART GETS DAMAGED IT CAN CAUSE A BIT OF SMOKE AND ELECTRICAL SMELL. THE ISSUE WAS LOCALLY SOLVED BY THE REPLACEMENT OF THE PCB ABSORBANCE MODULE AND THE ABSORBANCE LAMP.

Description of Event or Problem · 1

THE CUSTOMER HAD JUST STARTED THE INTEGRA 400+ (I400+) TO BEGIN THE MORNING CALIBRATION AND QC. THE INSTRUMENT WAS INITIALIZING WHEN THE CUSTOMER NOTICED A STRONG BURNING SMELL. THE CUSTOMER STOPPED THE ANALYZER AND CHECKED FOR THE SOURCE OF THE BURNING SMELL, AND INDICATED IT WAS COMING FROM THE LAMP AREA. THE CUSTOMER TRIED TO REPLACE THE LAMP, BUT WAS UNABLE TO REMOVE THE PLUG FROM THE OLD LAMP AS IT WAS MELTED INTO THE CONNECTOR. THERE WERE NO INSTRUMENT ALARMS. THE CUSTOMER INDICATED THERE WAS NO SMOKE, NO VISIBLE SPARKS AND NO FLAMES. THE LABORATORY WAS NOT EVACUATED DUE TO THE BURNING SMELL AND NO ONE HAD TO SEEK MEDICAL ATTENTION. THE FIELD SERVICE REPRESENTATIVE FOUND AN ISSUE WITH THE ABSORBANCE PRINTED CIRCUIT BOARD MODULE. HE REPLACED THE BOARD AND PERFORMED QC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
405007 COBAS INTEGRA 400 PLUS CLINICAL CHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS NA NA

Patients

Seq Age Sex Outcome Treatment
1