FDA Adverse Event Injury Summary report: N

PINNACLE MTL INS NEUT36IDX52OD

MDR report key: 3301152 · Received August 21, 2013

Report

Report Number
1818910-2013-24673
Event Type
Injury
Date Received
August 21, 2013
Date of Event
August 7, 2013
Report Date
August 7, 2013
Manufacturer
8010379 DEPUY INTL., LTD.
Product Code
KWA
PMA / PMN Number
PK083642
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A PREVIOUS REVIEW OF THE DEVICE HISTORY RECORDS FOR 2982992 LOT CODE DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE REMAINING PRODUCT AND LOT COMBINATIONS. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. D. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS HIGH COCR LEVELS. METALOSIS WAS FOUND IN AND AROUND THE JOINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404229 PINNACLE MTL INS NEUT36IDX52OD ACETABULAR LINER KWA 8010379 DEPUY INTL., LTD. 2982992

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention