SECUR-FIT HA PSL CUP/CLUSTR SHELL 54MM
Report
- Report Number
- 0002249697-2013-02752
- Event Type
- Injury
- Date Received
- August 16, 2013
- Date of Event
- June 1, 2013
- Report Date
- July 31, 2013
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- MEH
- PMA / PMN Number
- K942900
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ADDITIONAL INFORMATION (INCLUDING X-RAYS AND MEDICAL RECORDS) HAS BEEN REQUESTED. WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
AN EVENT REGARDING CORROSION INVOLVING A SECURFIT SHELL WAS REPORTED. BASED ON THE PROVIDED INFORMATION IT HAS BEEN DETERMINED THAT THIS EVENT IS ASSOCIATED WITH AN OFF-LABEL APPLICATION OF THE SECURFIT SHELL AS IT WAS USED WITH ANOTHER MANUFACTURER'S FEMORAL STEM AND HEAD. A REVIEW OF THE IFU CURRENTLY INCLUDED WITH THE DEVICE AT THE TIME OF MANUFACTURE, QIN 4302, STATES THE FOLLOWING WARNINGS: HOWMEDICA OSTEONICS CORP. STRONGLY ADVISES AGAINST THE USE OF ANOTHER MANUFACTURER¿S FEMORAL STEM COMPONENT WITH ANY HOWMEDICA OSTEONICS ACETABULAR SYSTEM COMPONENT. ANY SUCH USE WILL NEGATE THE RESPONSIBILITY OF HOWMEDICA OSTEONICS CORP. FOR THE PERFORMANCE OF THE RESULTING MIXED COMPONENT IMPLANT.
(B)(4) REPORT RECEIVED WITH FOLLOWING DESCRIPTION : "CHRONIC PAIN IN RELATION TO A MAJOR OSTEOLYSIS IN THE HIP. METALLOSIS PROBLEM " IT CONCERNS A PROTHEOS IMPLANT (STEM AND HEAD) WITH A STRYKER SECURIFIT CUP.
(B)(6) REPORT RECEIVED WITH FOLLOWING DESCRIPTION : "CHRONIC PAIN IN RELATION TO A MAJOR OSTEOLYSIS IN THE HIP. METALLOSIS PROBLEM." IT CONCERNS A PROTHEOS IMPLANT (STEM AND HEAD) WITH A STRYKER SECURIFIT CUP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 396873 | SECUR-FIT HA PSL CUP/CLUSTR SHELL 54MM | IMPLANT | MEH | STRYKER ORTHOPAEDICS-MAHWAH | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |