FDA Adverse Event
Malfunction
Summary report: N
QUANTUM SURGICAL TABLE
MDR report key: 32909
·
Received May 10, 1996
Report
- Report Number
- MW4001186
- Event Type
- Malfunction
- Date Received
- May 10, 1996
- Report Date
- July 17, 1995
- Manufacturer
- AMERICAN STERILIZER CO.
- Product Code
- FQO
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- HI, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
MECHANICAL FAILURE. POSSIBLE INJURY OR DEATH TO PT WHEN EQUIPMENT IN USE. THE UNIT'S MOTOR CONTROL ASSEMBLE FAILED DURING USE, CAUSING THE UNIT TO LOOSE CONTROL OF THE TABLE TOP, WHICH COULD RESULT IN THE PT BEING DROPPED WHEN IN USE. A LOCKPLATE FASTENER THAT SECURES THE TILT CYLINDER TO THE CYLINDER KNUCKLE WAS LOOSE. EXAMINATION OF THE LOCKING MECHANISM REVEALED THE HEX SCREWS COMPLETELY LOOSE ALLOWING THE CYLINDER TO UNSCREW ITSELF FROM THE CYLINDER KNUCKLE. THE MOTOR CONTROL ASSEMBLY CONSISTING OF THE TILT CYLINDER ASSEMBLY IS THE PROBLEM AREA. A LOOSE LOCK PLATE FASTENER THAT SECURES THE TILT CYLINDER TO THE CYLINDER KNUCKLE IS AT FAULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | QUANTUM SURGICAL TABLE | SURGICAL TABLE | FQO | AMERICAN STERILIZER CO. | 30805P |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO |