FDA Adverse Event
Injury
Summary report: N
HILL-ROM AIR SHIELDS
MDR report key: 328808
·
Received April 24, 2001
Report
- Report Number
- 328808
- Event Type
- Injury
- Date Received
- April 24, 2001
- Date of Event
- March 15, 2001
- Report Date
- April 3, 2001
- Manufacturer
- HILL-ROM AIRSHIELDS
- Product Code
- FMT
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
HILL ROM AIRSHIELD WARMER APPEARED TO BE HEATING BUT WASN'T. BABY'S TEMP WENT FROM 36 DEGREES CENTIGRADE TO 33 DEGREES CENTIGRADE. FAULTY HEATING ELEMENT WAS REPLACED. NO LONG TERM HARM TO BABY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 18832 | HILL-ROM AIR SHIELDS | WARMER | FMT | HILL-ROM AIRSHIELDS | 781 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 DAY | Required Intervention |