FDA Adverse Event Malfunction Summary report: N

XIA 3 TITANIUM ROD DIAM 6MM CP TI L 600MM

MDR report key: 3286502 · Received August 14, 2013

Report

Report Number
0009617544-2013-00310
Event Type
Malfunction
Date Received
August 14, 2013
Date of Event
July 16, 2013
Report Date
July 16, 2013
Manufacturer
STRYKER SPINE-FRANCE
Product Code
MNH
PMA / PMN Number
K071373
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF MADE AVAILABLE WILL BE REPORTED IN A SUPPLEMENTAL REPORT. METHOD, RESULT, AND CONCLUSION WILL BE MADE AVAILABLE FOLLOWING AN ENGINEERING EVALUATION.

Additional Manufacturer Narrative · 1

METHOD: DEVICE HISTORY REVIEW, RISK ASSESSMENT. RESULTS: THE CUSTOMER EVENT OF A XIA 3 TITANIUM ROD FRACTURE WAS CONFIRMED VIA VISUAL INSPECTION OF THE RETURNED DEVICE. THE IFU STATES THAT THE IMPLANT CAN BREAK OR BECOME DAMAGED AS A RESULT OF STRENUOUS ACTIVITY OR TRAUMA, AND THAT THE DEVICE MAY NEED TO BE REPLACED IN THE FUTURE. IT ALSO STATES THAT THE STRESSES AND STRAINS ON THE IMPLANTS MAY CAUSE METAL FATIGUE OR FRACTURE OR DEFORMATION OF THE IMPLANTS, BEFORE THE BONE GRAFT HAS BECOME COMPLETELY CONSOLIDATED. THE EXTENDED PERIOD OF TIME BETWEEN SURGERIES (~ 18 MONTHS) AND THE EVENT SUGGESTS THAT FUSION OF THE SPINE MAY NOT HAVE OCCURRED AND AS A RESULT THE CONSTRUCT WAS LEFT TO SUPPORT THE ENTIRE WEIGHT OF THE SPINE, WHICH MAY HAVE LED TO A POSSIBLE BREAKAGE. CONCLUSION: THE ROOT CAUSE OF THE FAILURE IS LIKELY A FATIGUE FRACTURE DUE TO PROLONGED IMPLANTATION WITHIN THE BODY WITH A POSSIBLE LACK OF FUSION CONTRIBUTING TO THE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT "AFTER THE PRIMARY SURGERY, THE ROD BREAKAGE WAS CONFIRMED."

Description of Event or Problem · 1

IT WAS REPORTED THAT "AFTER THE PRIMARY SURGERY, THE ROD BREAKAGE WAS CONFIRMED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
390439 XIA 3 TITANIUM ROD DIAM 6MM CP TI L 600MM IMPLANT MNH STRYKER SPINE-FRANCE UNK

Patients

Seq Age Sex Outcome Treatment
1 75 YR