SCRDRIVER SHAFT 1.5 W/HOLD-SLEEVE L66
Report
- Report Number
- 1719045-2013-01882
- Event Type
- Malfunction
- Date Received
- August 14, 2013
- Date of Event
- July 18, 2013
- Report Date
- July 18, 2013
- Manufacturer
- SYNTHES MONUMENT
- Product Code
- HXX
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN SINCE THE EVALUATION OF THIS DEVICE IS ONGOING. THE MANUFACTURING RECORDS WERE REQUESTED SINCE THE LOT NUMBER WAS PROVIDED. PLACEHOLDER.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED THAT SYNTHES MONUMENT ASSEMBLED THE STARDRIVE SCREWDRIVER SHAFT W/HOLDING SLEEVE/T4/66MM, PART NUMBER 03.114.009, LOT NUMBER 7054679. THERE WERE NO MRRS, NCRS, OR COMPLAINT-RELATED ISSUES WITH THIS LOT. 25 PARTS WERE RELEASED TO THE WAREHOUSE ON OCTOBER 09, 2012.
AN INVESTIGATION SHOWED THAT THE TIP IS BROKEN OFF. THE REVIEW OF THE PRODUCTION HISTORY REVEALED THAT THIS INSTRUMENT IS MANUFACTURED ACCORDING TO THE SPECIFICATIONS. NO MANUFACTURING RELATED ISSUES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT WERE FOUND. BASED ON THE FINDINGS WE HAVE TO ASSUME THAT HIGH APPLIED MECHANICAL FORCE DURING SCREW TIGHTENING CAUSED THE BREAKAGE.
ON (B)(6) 2013, WHILE INSERTING A LOCKING SCREW (LCP COHA SET) THE TIP OF SCREWDRIVER BROKE. PART 03.114.009 CONCERNS AN ASSEMBLY GROUP, BUT ONLY SCREWDRIVER (PART 03.114.010, LOT 7010083) WAS SENT FOR INVESTIGATION, THE HOLDING SLEEVE WAS NOT AVAILABLE. THIS IS 1 OF 1 DEVICES FOR THIS EVENT REPORTED ON COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390141 | SCRDRIVER SHAFT 1.5 W/HOLD-SLEEVE L66 | HXX | SYNTHES MONUMENT | 7054679 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |