FDA Adverse Event Injury Summary report: N

PFC*SIGMA C/S NPOR FEM RT SZ 4

MDR report key: 3267851 · Received August 7, 2013

Report

Report Number
1818910-2013-23187
Event Type
Injury
Date Received
August 7, 2013
Date of Event
August 8, 2008
Report Date
December 20, 2013
Manufacturer
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT R
Product Code
JWH
PMA / PMN Number
PK950010
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED AND IS PRESUMED YET IMPLANTED, AS NO REVISION SURGERY HAS BEEN REPORTED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED AND ARE PRESUMED YET IMPLANTED. REVIEW OF THE SUPPLIED PATIENT MEDICAL RECORDS CONFIRMED PATIENT LESION/SCAR TISSUE THAT COULD CONTRIBUTE TO THE REPORTED PATELLAR GRINDING. NO EVIDENCE WAS FOUND SUGGESTING PRODUCT ERROR WAS A CONTRIBUTING FACTOR TO THE PATELLAR GRINDING AND THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

CLINICAL REPORT STATES THE PATIENT HAD PATELLAR GRIND SYNDROME. THE PATIENT UNDERWENT AN ARTHROSCOPY FOR TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
370395 PFC*SIGMA C/S NPOR FEM RT SZ 4 FEMORAL JWH THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT R 147054AB

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention