CRYO-SPRAY ABLATION SYSTEM
Report
- Report Number
- 3004534508-2013-00002
- Event Type
- Injury
- Date Received
- August 1, 2013
- Date of Event
- July 2, 2013
- Report Date
- August 1, 2013
- Manufacturer
- CSA MEDICAL
- Product Code
- GEH
- PMA / PMN Number
- K072651
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE ON SITE EVALUATION: CSA MEDICAL SVC PERSONNEL WAS SENT TO THE SITE ON (B)(4) 2013 (POST EVENT), TO PERFORM FIELD VERIFICATION TESTING PER WORK INSTRUCTION (WI). THE CONSOLE ITSELF (CC2-NAM) PASSED ALL TESTS, AND WAS FOUND TO BE WITHIN SPECIFICATION. TESTING INITIALLY SHOWED THAT THE MEASURED VACUUM SOURCE USING A CSA MEDICAL RECOMMENDED PUMP WAS BELOW SPECIFICATION. IT SHOULD BE NOTED THAT THE VACUUM SOURCE (SUCTION PUMP AND HOSE) WAS TESTED SEPARATE FROM THE CONSOLE, PER WI. SVC FOUND THAT THOSE BETWEEN CONSOLE AND SUCTION PUMP CONSISTED OF 2 SECTIONS INAPPROPRIATELY CONNECTED TOGETHER WITH A SMALL DIAMETER COUPLING. SVC REMOVED THE COUPLING AND ONE HOSE SECTION AND RETESTED THE VACUUM WITH THE MAIN HOSE SECTION IN PLACE. RESULTS EXCEEDED MINIMUM SPECIFICATION. THE FIELD VERIFICATION TESTING WAS THEN COMPLETED AND ALL TESTS PASSED. THE SITE WAS INFORMED OF TEST RESULTS. NOTE: A RELATIONSHIP BETWEEN REDUCED SUCTION AT THE TIME OF FIELD TESTING AND THE ADVERSE EVENT COULD NOT BE DETERMINED.
DURING CRYOTHERAPY OF A SITE IN THE DISTAL 3RD OF THE STOMACH, PT'S ABDOMEN BECAME DISTENDED. TREATMENT WAS STOPPED. SUBSEQUENTLY, PT DEVELOPED ABDOMINAL PAIN. X-RAY SHOWED PNEUMOPERITONEUM. PT WAS HOSPITALIZED, TREATED WITH BOWEL REST AND ANTIBIOTICS, AND WAS DOING WELL WITHOUT SURGERY. DOCTOR PERFORMED PARACENTESIS 2X TO RELEASE ACCUMULATED INTRA ABDOMINAL GAS. PT WAS RELEASED FROM HOSP ON (B)(6) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359920 | CRYO-SPRAY ABLATION SYSTEM | CRYOSPRAY | GEH | CSA MEDICAL | CC2-NAM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Hospitalization| R |