COBRA GRASPER INSTRUMENT
Report
- Report Number
- 2955842-2013-02885
- Event Type
- Malfunction
- Date Received
- August 1, 2013
- Date of Event
- June 28, 2013
- Report Date
- November 22, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND DID NOT EXHIBIT DAMAGE. THE CABLE SEGMENT STUCK OUT AT WRIST. THE CABLE BROKE UNDER TENSILE LOADING. THE OTHER CABLES AT WRIST WERE UNDAMAGED. NO OTHER DAMAGE FOUND.
THE INSTRUMENT AND/OR ACCESSORY HAVE NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION), OR IF ADDITIONAL INFORMATION IS RECEIVED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE THE COBRA GRASPER INSTRUMENT CABLE BROKE AT THE WRIST. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360431 | COBRA GRASPER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420190-07 | M20111110 426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES |