FDA Adverse Event Malfunction Summary report: N

EXTENSION SET

MDR report key: 3260130 · Received August 1, 2013

Report

Report Number
1416980-2013-20385
Event Type
Malfunction
Date Received
August 1, 2013
Date of Event
July 8, 2013
Report Date
July 8, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K925403
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT IS FOR SYSTEM ERROR 2240, WHERE THE HOME PATIENT (HP) CONNECTED PATIENT LINE EXTENSION SET AFTER THE PRIMING CYCLE OF THE THERAPY WAS COMPLETED. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN "THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE", WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. IT INSTRUCTS THE USER TO CONNECT THE PATIENT LINE EXTENSION TO THE PATIENT LINE PRIOR TO PRIMING. ALSO, IT INSTRUCTS THE USER TO MAKE SURE THE PATIENT LINE EXTENSION IS CORRECTLY POSITIONED IN THE ORGANIZER AND VERIFY THAT THE PATIENT LINE EXTENSION IS PLACED IN THE LEFT SLOT OF THE BLUE ORGANIZER. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ANY RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HOME PATIENT (HP) HAD A SYSTEM ERROR 2240 (AIR IN TUBING) ON THE HOMECHOICE (HC) DURING THE INITIAL DRAIN, WHILE THE HP WAS CONNECTED. THE HP WAS USING EXTENSION LINE BUT DID NOT CONNECT IT TO THE PATIENT LINE PRIOR TO PRIMING. AS A RESULT, THE PATIENT LINE EXTENSION SET DID NOT GET PRIMED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND ITS CAUSE. THE TSR HAD THE HP CLOSE ALL THE CLAMPS AND CYCLE THE POWER IN ORDER TO CLEAR THE ALARM. THE TSR ADVISED THE HP TO START THE THERAPY OVER. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
360547 EXTENSION SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 66 YR HOMECHOICE| HOMECHOICE AUTOMATED PD SET WITH CASSETTE