EXTENSION SET
Report
- Report Number
- 1416980-2013-20385
- Event Type
- Malfunction
- Date Received
- August 1, 2013
- Date of Event
- July 8, 2013
- Report Date
- July 8, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- PMA / PMN Number
- K925403
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS REPORT IS FOR SYSTEM ERROR 2240, WHERE THE HOME PATIENT (HP) CONNECTED PATIENT LINE EXTENSION SET AFTER THE PRIMING CYCLE OF THE THERAPY WAS COMPLETED. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN "THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE", WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. IT INSTRUCTS THE USER TO CONNECT THE PATIENT LINE EXTENSION TO THE PATIENT LINE PRIOR TO PRIMING. ALSO, IT INSTRUCTS THE USER TO MAKE SURE THE PATIENT LINE EXTENSION IS CORRECTLY POSITIONED IN THE ORGANIZER AND VERIFY THAT THE PATIENT LINE EXTENSION IS PLACED IN THE LEFT SLOT OF THE BLUE ORGANIZER. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ANY RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE HOME PATIENT (HP) HAD A SYSTEM ERROR 2240 (AIR IN TUBING) ON THE HOMECHOICE (HC) DURING THE INITIAL DRAIN, WHILE THE HP WAS CONNECTED. THE HP WAS USING EXTENSION LINE BUT DID NOT CONNECT IT TO THE PATIENT LINE PRIOR TO PRIMING. AS A RESULT, THE PATIENT LINE EXTENSION SET DID NOT GET PRIMED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND ITS CAUSE. THE TSR HAD THE HP CLOSE ALL THE CLAMPS AND CYCLE THE POWER IN ORDER TO CLEAR THE ALARM. THE TSR ADVISED THE HP TO START THE THERAPY OVER. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360547 | EXTENSION SET | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | HOMECHOICE| HOMECHOICE AUTOMATED PD SET WITH CASSETTE |