FDA Adverse Event
Other
Summary report: N
HILL-ROM
MDR report key: 3260007
·
Received July 26, 2013
Report
- Report Number
- MW5031240
- Event Type
- Other
- Date Received
- July 26, 2013
- Report Date
- July 26, 2013
- Manufacturer
- HILL ROM
- Product Code
- FNL
- Report Source
- Voluntary report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
(B)(6) WAS EXPERIENCED REPEATED CLOSE CALLS RELATED TO THE FUNCTION OF THE HILL ROM EXCEL BARIATRIC BEDS. WE RENT BOTH THE EXCEL AIR LOSS MATTRESS AND THE EXCEL FOAM MATTRESS. IN ORDER TO FIT THE BED INTO OUR BARIATRIC ROOMS, WE MUST DEFLATE THE MATTRESS AND COLLAPSE THE EDGES OF THE BED DECK, WHICH MAKES THE BED NARROWER. WHEN SETTING UP THE BED, THE MATTRESS CAN BE RE-INFLATED WITHOUT EXTENDING THE BED DECK. THIS PRESENTS A FALL RISK WHEN A PT IS GETTING IN OR OUT OF BED AND THERE IS NO DECK UNDERNEATH THE EDGE OF THE MATTRESS. WE RECOMMEND ENGINEERING THE BEDS SO THAT THE BED DECK MUST BE EXTENDED BEFORE THE MATTRESS CAN BE INFLATED.