FDA Adverse Event Other Summary report: N

TANDEM T: SLIM INSULIN DELIVERY SYSTEM

MDR report key: 3254150 · Received July 24, 2013

Report

Report Number
3007981285-2013-00039
Event Type
Other
Date Received
July 24, 2013
Date of Event
July 8, 2013
Report Date
July 8, 2013
Manufacturer
TANDEM DIABETES CARE INC.
Product Code
LZG
PMA / PMN Number
K111210
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING FURTHER INVESTIGATION PATIENT INFORMED TANDEM THAT UNIT HAD PROVIDED THREE OCCLUSION ALARMS. PATIENT CLEARED THE FIRST TWO ALARMS AND APPARENTLY DID NOT HEAR THE THIRD ALARM. PATIENT WOKE UP WITH A BG OF 400 AND CHANGED OUT HER INFUSION SET AND CARTRIDGE. PATIENT SAW BLOOD IN THE CANNULA AND WAS BLEEDING FROM HER SITE INDICATING UNIT WORKED AS INTENDED. PATIENT WAS SEEN BY HER HEALTH CARE PROVIDER (HCP) WHO SENT HER TO THE HOSPITAL, WHERE THEY WERE ABLE TO GET HER BG LEVELS BACK WITHIN HER TARGET RANGE. USER GUIDE STATES; IF YOUR T: SLIM PUMP DETECTS INSULIN DELIVERY IS BLOCKED, THE OCCLUSION ALARM OCCURS AND ALL INSULIN DELIVERIES ARE STOPPED. CHECK THE CARTRIDGE, TUBING AND CANNULA TO HELP DETERMINE THE CAUSE. EVENT NOT LIKELY TO LEAD TO DEATH, T: SLIM USER GUIDE INSTRUCTS USERS TO "ROUTINELY CHECK THEIR BG LEVELS AT LEAST 4 TIMES DAILY (OPTIMALLY 6-8 TIMES DAILY) IN ORDER TO DETECT HYPERGLYCEMIA AND HYPOGLYCEMIA EARLY." ADDITIONALLY, T: SLIM USER GUIDE CAUTIONS USERS "TO PREVENT DKA OR A VERY HIGH BG, YOU MUST BE PREPARED TO INJECT INSULIN IF DELIVERY IS INTERRUPTED FOR ANY REASON." IT IS COMMON PRACTICE FOR INSULIN USERS TO CHECK BLOOD GLUCOSE VALUES FREQUENTLY TO PREVENT LEVELS FROM BECOMING SEVERE ENOUGH TO CAUSE SERIOUS COMPLICATIONS.

Description of Event or Problem · 1

RECEIVED INFO FROM A TANDEMS SALES REP INDICATING A PATIENT WAS TAKEN TO THE HOSPITAL FOR HIGH BG LEVELS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344547 TANDEM T: SLIM INSULIN DELIVERY SYSTEM INSULIN PUMP LZG TANDEM DIABETES CARE INC. 004628

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization