FDA Adverse Event Injury Summary report: N

M2A 38MM MODULAR HEAD+6MM NK NO SKRT

MDR report key: 3253871 · Received July 30, 2013

Report

Report Number
0001825034-2013-02963
Event Type
Injury
Date Received
July 30, 2013
Date of Event
July 1, 2013
Report Date
October 2, 2013
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK011110
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DIMENSIONAL EVALUATION FOUND COMPONENT TO BE WITHIN APPROPRIATE DESIGN SPECIFICATION.

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 15 STATES, ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PATIENT¿S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6)2005. SUBSEQUENTLY, A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2013, DUE TO ALLEGED ELEVATED METAL ION LEVELS. A REVIEW OF INVOICE HISTORY CONFIRMED THE MODULAR HEAD WAS REMOVED AND REPLACED WITH MODULAR HEAD AND POLYETHYLENE LINER. NO FURTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
356509 M2A 38MM MODULAR HEAD+6MM NK NO SKRT PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 583740

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R