FDA Adverse Event Malfunction Summary report: N

SITTER II

MDR report key: 3253646 · Received July 17, 2013

Report

Report Number
2020362-2013-00321
Event Type
Malfunction
Date Received
July 17, 2013
Report Date
May 10, 2013
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL, RESULTS: EVAL OF THE RETURNED PRODUCT DID NOT CONFIRM THE REPORTED ISSUE; THE POWER AND HOLD BUTTONS ARE NOT MISSING BUT DO HAVE SIGNS OF DAMAGE TO THEM. HOWEVER, THE ALARM POWERS ON AND WHILE TESTING THE LOW BATTERY FUNCTION, THE ALARM LOST POWER. THE LOW BATTERY LED LIGHT IS FAINT. RE-INSTALLED THE BATTERIES AND THE UNIT HAS POWER BUT THE TONE AND VOICE HAS A STATIC SOUND. THE SCREWS ON THE BACK OF THE ALARM CASE HAVE RUST ON THEM. MORE INFO RECEIVED IS THE VOLTAGE REGULATOR IS LOW. NOTE: THE INSTRUCTIONS FOR USE STATE: THE LOW BATTERY LIGHT WILL FLASH RED WHEN NEW BATTERIES ARE NEEDED. CHANGE BATTERIES AT ONCE. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THE POWER BUTTON IS MISSING ON THE ALARM. CUSTOMER DID NOT PROVIDE A DATE WHEN THIS WAS DISCOVERED. NO PT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331802 SITTER II KMI J. T. POSEY CO. 8281 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK LOT # UNK| OVER-THE-MATTRESS SENSOR PAD: MODEL 8307,