FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3253312 · Received July 29, 2013

Report

Report Number
3004209178-2013-96199
Event Type
Injury
Date Received
July 29, 2013
Date of Event
April 10, 2013
Report Date
July 4, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

ALL BUTTONS FUNCTION PROPERLY. NO BUTTON ERROR ALARMS NOTED. HOWEVER, MOISTURE DAMAGE WAS NOTED ON KEYPAD TRACES DURING VISUAL INSPECTION. UNABLE TO PRIME DURING PRIME TEST DUE TO MOISTURE DAMAGE NOTED IN FORCE SENSOR. UNABLE TO PERFORM OCCLUSION AND EXCESSIVE NO DELIVERY ALARM TEST DUE TO PRIME ANOMALY. THE INSULIN PUMP PASSED DISPLACEMENT TEST. MISSING END CAP STICKER NOTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT SUDDENLY HER INSULIN PUMP STOPPED WORKING, AND THE BUTTONS WERE UNRESPONSIVE. THE CUSTOMER WAS EXPERIENCING HIGH BLOOD GLUCOSE OVER 410MG/DL. THE CUSTOMER MENTIONED BEING OFF THE INSULIN PUMP FOR TWO HOURS WITHOUT TREATING AFTER EATING. DURING THE CALL, THE CUSTOMER MENTIONED BEING HOSPITALIZED DUE TO LOW BLOOD GLUCOSE OF 30S OR 40S MG/DL ON (B)(6), 2013. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER WAS UNABLE TO DESCRIBE ANY POSSIBLE DAMAGE TO THE DRIVE SUPPORT CAP. THE CUSTOMER BELIEVES THAT SHE HAD OVER CORRECTED AND DID NOT EAT ENOUGH TO COVER THE BOLUS. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353965 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization