FDA Adverse Event
Injury
Summary report: N
PARADIGM INSULIN INFUSION PUMP
MDR report key: 3253309
·
Received July 29, 2013
Report
- Report Number
- 3004209178-2013-96173
- Event Type
- Injury
- Date Received
- July 29, 2013
- Date of Event
- July 25, 2013
- Report Date
- July 25, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LZG
- PMA / PMN Number
- K040676
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED ALL FUNCTIONAL TESTS, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, EXCESSIVE NO DELIVERY ALARM TEST.
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED WITH HIGH BLOOD GLUCOSE OF 628MG/DL. TROUBLESHOOTING COULD NOT BE PERFORMED, AND THE INSULIN PUMP ALARMED NO DELIVERY. WHEN THE CUSTOMER CHANGED THE INFUSION SET THE CANNULA WAS BENT. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 353964 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-715WWS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |