FDA Adverse Event Malfunction Summary report: N

COULTER® ACT DIFF ANALYZER

MDR report key: 3253135 · Received July 29, 2013

Report

Report Number
1061932-2013-01593
Event Type
Malfunction
Date Received
July 29, 2013
Date of Event
July 10, 2013
Report Date
July 10, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K973634
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE ON (B)(4) 2013. THE FSE OBSERVED A PROBE LEAK AND FOUND THAT DILUENT FILTERS HAD NOT BEEN CHANGED SINCE DECEMBER 2012. THE FSE REPLACED DILUENT FILTERS INSIDE THE RIGHT HAND DOOR AND 3 HYDRO BLUE FILTERS. THE FSE RAN SEVERAL STARTUP'S AND NO LEAK WAS PRESENT. WHILE ON SITE, THE FSE ALSO CLEANED THE PROBE WIPE AS PREVENTIVE MAINTENANCE. THE FSE ADVISED CUSTOMER TO INCORPORATE INTO REGULAR MAINTENANCE. FAILURE MODE OF THE LEAK WAS RELATED TO THE DILUENT FILTERS. BECKMAN INTERNAL IDENTIFIER NUMBER FOR THIS REPORT IS (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT APPROXIMATELY 1 ML OF FLUID WAS LEAKING FROM THE PROBE IN THE COULTER AC*T-DIFF ANALYZER. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED BY THE CUSTOMER. THERE IS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN IN PLACE AT THE FACILITY. THE OPERATOR WAS WEARING A LABORATORY COAT AND GLOVES AT THE TIME OF THE INCIDENT. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. THERE WAS NO IMPACT TO PATIENT RESULTS. NO ERRONEOUS RESULTS WERE GENERATED AS A RESULT OF THIS EVENT. NO DEATH OR INJURY IS ATTRIBUTED TO THIS EVENT AND THERE WAS NO CHANGE TO PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353450 COULTER® ACT DIFF ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA

Patients

Seq Age Sex Outcome Treatment
1