FDA Adverse Event
Malfunction
Summary report: N
FENESTRATED BIPOLAR FORCEPS INSTRUMENT
MDR report key: 3253121
·
Received July 29, 2013
Report
- Report Number
- 2955842-2013-02810
- Event Type
- Malfunction
- Date Received
- July 29, 2013
- Date of Event
- June 19, 2013
- Report Date
- July 2, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED A BROKEN PITCH CABLE AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP REMAINED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT EXCESSIVE DAMAGE. NO OTHER DAMAGE WAS FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE USER FACILITY IDENTIFIED WIRES POKING OUTSIDE NEAR THE TIP OF THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 354064 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M10130416 667 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS, ACCESSORIES, ESU |