FDA Adverse Event
Malfunction
Summary report: N
SECURE II MED/SURG BED
MDR report key: 3253074
·
Received July 29, 2013
Report
- Report Number
- 0001831750-2013-06755
- Event Type
- Malfunction
- Date Received
- July 29, 2013
- Date of Event
- July 2, 2013
- Report Date
- July 2, 2013
- Manufacturer
- STRYKER MEDICAL-KALAMAZOO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
SUPPLEMENTAL SUBMITTED AS THE INVESTIGATION CONCLUDED THAT THE CPU AND POWER SUPPLY BOARD WERE BURNED. THERE WAS NOTHING WRONG WITH POWER BOARD AS REPORTED.
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THERE WAS NO POWER TO THE BED DUE TO MALFUNCTION CPU AND POWER SUPPLY BOARD WERE BURNED. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THERE WAS NO POWER TO THE BED DUE TO MALFUNCTION POWER SUPPLY, POWER BOARD AND CPU BOARD. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 353030 | SECURE II MED/SURG BED | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | STRYKER MEDICAL-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |