FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3253015 · Received July 29, 2013

Report

Report Number
2032227-2013-03190
Event Type
Malfunction
Date Received
July 29, 2013
Date of Event
July 2, 2013
Report Date
July 2, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURN FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS. THE INSULIN PUMP WAS RECEIVED WITH NORMAL OPERATING CURRENTS AND NO UNEXPECTED OFF NO POWER OR LOW BATTERY ALARM NOTED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP DRIVE SUPPORT CAP IS CRACK. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352447 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAH

Patients

Seq Age Sex Outcome Treatment
1 49 YR