FDA Adverse Event Malfunction Summary report: N

12.4CM HEAVY DUTY LONG ATTCH

MDR report key: 3252957 · Received July 29, 2013

Report

Report Number
1045834-2013-03721
Event Type
Malfunction
Date Received
July 29, 2013
Date of Event
September 28, 2011
Report Date
October 6, 2011
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE REPORTED CONDITION OF OVERHEATING WAS CONFIRMED. (B)(4) EVALUATED THE DEVICE, AND THE REPORTED HEAT WAS CONFIRMED; THE ATTACHMENT WAS TESTED AND FOUND TO EXCEED TEMPERATURE SPECIFICATIONS. IT WAS CONCLUDED THAT THE HEAT WAS CAUSED BY WORN/DAMAGED BEARINGS. THE DEVICE WAS DISASSEMBLED, AND DRIED BIOLOGICAL DEBRIS WAS OBSERVED COATING THE BEARINGS OF THE ATTACHMENTS. IT'S LIKELY THE FRICTION CREATED BY THE PRESENCE OF THE DEBRIS CONTRIBUTED TO THE REPORTED HEAT. THIS CONDITION IS INDICATIVE OF IMPROPER OR INEFFECTIVE CLEANING AND MAINTENANCE OF THE DEVICES. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).

Description of Event or Problem · 1

REPORT RECEIVED FROM THE USA STATING THAT THE DEVICE WAS OVERHEATING. THE DEVICE WAS NOT BEING USED DURING SURGERY. IT IS UNKNOWN IF INJURY OR MEDICAL INTERVENTION OCCURRED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352255 12.4CM HEAVY DUTY LONG ATTCH MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1