12.4CM HEAVY DUTY LONG ATTCH
Report
- Report Number
- 1045834-2013-03721
- Event Type
- Malfunction
- Date Received
- July 29, 2013
- Date of Event
- September 28, 2011
- Report Date
- October 6, 2011
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE REPORTED CONDITION OF OVERHEATING WAS CONFIRMED. (B)(4) EVALUATED THE DEVICE, AND THE REPORTED HEAT WAS CONFIRMED; THE ATTACHMENT WAS TESTED AND FOUND TO EXCEED TEMPERATURE SPECIFICATIONS. IT WAS CONCLUDED THAT THE HEAT WAS CAUSED BY WORN/DAMAGED BEARINGS. THE DEVICE WAS DISASSEMBLED, AND DRIED BIOLOGICAL DEBRIS WAS OBSERVED COATING THE BEARINGS OF THE ATTACHMENTS. IT'S LIKELY THE FRICTION CREATED BY THE PRESENCE OF THE DEBRIS CONTRIBUTED TO THE REPORTED HEAT. THIS CONDITION IS INDICATIVE OF IMPROPER OR INEFFECTIVE CLEANING AND MAINTENANCE OF THE DEVICES. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).
REPORT RECEIVED FROM THE USA STATING THAT THE DEVICE WAS OVERHEATING. THE DEVICE WAS NOT BEING USED DURING SURGERY. IT IS UNKNOWN IF INJURY OR MEDICAL INTERVENTION OCCURRED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352255 | 12.4CM HEAVY DUTY LONG ATTCH | MOTOR, DRILL, ELECTRIC - ATTACHMENT | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |