FDA Adverse Event Malfunction Summary report: N

8CM ANGLE ATTACHMENT

MDR report key: 3252953 · Received July 29, 2013

Report

Report Number
1045834-2013-03716
Event Type
Malfunction
Date Received
July 29, 2013
Date of Event
March 8, 2011
Report Date
March 14, 2011
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE REPORTED CONDITION OF "SPINS FREE" WAS NOT CONFIRMED - THE ATTACHMENT PERFORMED AS DESIGNED. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).

Description of Event or Problem · 1

REPORT RECEIVED FROM (B)(4) STATING THAT THE DEVICE "SPINS FREE". THE DEVICE WAS NOT BEING USED DURING SURGERY. IT IS UNKNOWN IF INJURY OR MEDICAL INTERVENTION OCCURRED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352128 8CM ANGLE ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1