FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3252847 · Received July 29, 2013

Report

Report Number
2531779-2013-12025
Event Type
Malfunction
Date Received
July 29, 2013
Report Date
July 1, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 10/07/2013 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/16/2013 WITH THE FOLLOWING FINDINGS: THE ¿LOW BATTERY¿ WARNING AND ¿REPLACE BATTERY¿ ALARM WAS NOTED IN THE BLACK BOX ON 06/30/2013. A ¿LOW BATTERY¿ WARNING AND A ¿REPLACE BATTERY¿ ALARMS WERE REPRODUCED FOR THE INVESTIGATION. BOTH OCCURRED AT THE CORRECT VOLTAGE. BOTH RECORDED IT THE CORRECT ORDER. A 10UNIT AUDIO AND 10UNIT NORMAL BOLUS WAS SUCCESSFULLY PERFORMED ON THE PUMP. BOTH WERE ACCURATELY RECORDED IN THE PUMP HISTORY. NO ALARMS OCCURRED DURING TESTING. THE REPORTED COMPLAINT WAS UNABLE TO BE DUPLICATED DURING TESTING.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS ISSUE. REPORTEDLY, THE PUMP EMITTED A LOW BATTERY ALARM AND REPLACE BATTERY ALARM AT THE SAME TIME AT 7:22 ON (B)(6) 2013. THE PUMP AGAIN ALARMED FOR LOW BATTERY SEVERAL HOURS LATER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353679 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 10 YR