FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3252752 · Received July 25, 2013

Report

Report Number
1720753-2013-08567
Event Type
Malfunction
Date Received
July 25, 2013
Date of Event
July 16, 2013
Report Date
July 25, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN OVER-THE-PHONE INVESTIGATION. THE CUSTOMER WAS INSTRUCTED TO RESET THE DISPLAY ADAPTOR BOARD SWITCH. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349280 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1