INLAY OPTIMA MULTILENGTH URETERAL STENT
Report
- Report Number
- 1018233-2013-02989
- Event Type
- Malfunction
- Date Received
- July 25, 2013
- Date of Event
- March 27, 2013
- Report Date
- May 29, 2013
- Manufacturer
- PRODUCTS PARA EL CUIDADO DELA SALUD
- Product Code
- FAD
- PMA / PMN Number
- K043193
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- UNKNOWN
Narratives
THE COMPLAINT IS NOT CONFIRMED. THE STENT WAS RECEIVED BROKEN IN TWO PIECES, BREAKING WHEN ATTEMPTING TO REMOVE THE GUIDEWIRE. DURING THE FUNCTIONAL EVALUATION OF THE RETURNED STENT, THE GUIDE WIRE ADVANCED OVER THE STENT WITHOUT ANY DIFFICULTY AND THE FAILURE COULD NOT BE DUPLICATED. THIS TEST WAS ALSO PERFORMED ON THE BROKEN PIGTAIL AND NO DISCREPANCIES WERE NOTICED. DIMENSIONS OF THE ID AND OD OF THE STENT WERE NOT TAKEN DUE TO STRESS MARKS ALONG THE SURFACE OF THE STENT AND PARTS DISPLAYED SLIGHT DEFORMATION. DIMENSIONS OF THE EYE PERFORATION WERE FOUND TO BE DEFORMED NEAR TO THE BREAKAGE SITE AND WERE FOUND TO BE OUT OF THE SPECIFICATIONS. THE DEFORMATION IS BELIEVED TO HAVE OCCURRED WHEN THE STENT WAS STRETCHED. DIRECTIONS FOR USE STATE: "DETERMINE THE PROPER STENT LENGTH FOR THE PATIENT. THIS IS GENERALLY CALCULATED FROM THE BASELINE PYELOGRAM. ACCURATE MEASUREMENTS WILL OPTIMIZE DRAINAGE EFFICIENCY AND PATIENT COMFORT. SUBMERGE STENT IN STERILE WATER TO ACTIVATE THE COATING. INSERT THE CYSTOSCOPE THEN PASS THE GUIDEWIRE THROUGH THE SCOPE UNTIL THE TIP IS IN THE RENAL PELVIS. MOVE THE PIGTAIL STRAIGHTENER OVER THE PROXIMAL END (KIDNEY COIL END) OF THE URETERAL STENT ALLOWING EASIER INSERTION ONTO THE GUIDEWIRE. REMOVE PIGTAIL STRAIGHTENER ONCE THE STENT IS SECURE ON THE GUIDEWIRE. PASS THE STENT OVER THE GUIDEWIRE THROUGH THE CYSTOSCOPE BY USING THE PUSH CATHETER FOR PROPER PLACEMENT. WATCH THE DISTAL END (BLADDER COIL END) OF THE STENT OR THE RADIOPAQUE, PROXIMAL END OF THE PUSHER. STOP ADVANCING WHEN THE STENT'S DISTAL END MARKER REACHES THE URETEROVESICAL JUNCTION (UVJ). (SEE BELOW FOR PROPER PLACEMENT DIRECTIONS ON THE MULTI-LENGTH URETERAL STENT). WITHDRAWAL THE GUIDEWIRE SLOWLY. THE STENT WILL FORM A PIGTAIL AUTOMATICALLY. CAREFULLY REMOVE THE PUSH CATHETER ACTIVATE THE GUIDEWIRE COATING ACCORDING TO THE "INSTRUCTIONS FOR USE" FOUND WITHIN THE GUIDEWIRE PACKAGING." (B)(4).
(B)(4). IT WAS REPORTED THAT THE STENT DID NOT PROGRESS OVER THE GUIDEWIRE. THE STENT WAS BROKEN WHEN THEY ATTEMPTED TO REMOVE THE GUIDEWIRE. THERE WAS PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 347995 | INLAY OPTIMA MULTILENGTH URETERAL STENT | FAD | PRODUCTS PARA EL CUIDADO DELA SALUD | NA | NGTE0316 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |