FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 3252507 · Received July 29, 2013

Report

Report Number
1416980-2013-20029
Event Type
Malfunction
Date Received
July 29, 2013
Date of Event
July 3, 2013
Report Date
July 4, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS AVAILABLE FOR ANALYSIS AND HAS BEEN RECEIVED FOR EVALUATION. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED FOR LOT NUMBER H13C07061 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION NOTED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). VISUAL INSPECTION AND CLAMP FUNCTION TESTING OF THE RETURNED SAMPLE IDENTIFIED THE BROKEN OCCLUDER FEET. THE REPORTED ISSUE WAS CONFIRMED. LEAK TESTING AND CLEAR PASSAGE TESTING WERE PERFORMED WITH NO ISSUES NOTED. THE CAUSE COULD NOT BE DETERMINED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BLUE PART (TWIST CLAMP) OF THE MINICAP TRANSFER SET HAD LEAKED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352385 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H13C07061

Patients

Seq Age Sex Outcome Treatment
1