MINICAP TRANSFER SET
Report
- Report Number
- 1416980-2013-20029
- Event Type
- Malfunction
- Date Received
- July 29, 2013
- Date of Event
- July 3, 2013
- Report Date
- July 4, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- PMA / PMN Number
- K882498
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE IS AVAILABLE FOR ANALYSIS AND HAS BEEN RECEIVED FOR EVALUATION. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED FOR LOT NUMBER H13C07061 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION NOTED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.
(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). VISUAL INSPECTION AND CLAMP FUNCTION TESTING OF THE RETURNED SAMPLE IDENTIFIED THE BROKEN OCCLUDER FEET. THE REPORTED ISSUE WAS CONFIRMED. LEAK TESTING AND CLEAR PASSAGE TESTING WERE PERFORMED WITH NO ISSUES NOTED. THE CAUSE COULD NOT BE DETERMINED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE BLUE PART (TWIST CLAMP) OF THE MINICAP TRANSFER SET HAD LEAKED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352385 | MINICAP TRANSFER SET | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME | H13C07061 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |