GYNECARE MORCELLEX TISSUE MORCELLATOR
Report
- Report Number
- 2210968-2013-14888
- Event Type
- Malfunction
- Date Received
- July 29, 2013
- Date of Event
- July 12, 2013
- Report Date
- July 12, 2013
- Manufacturer
- ETHICON INC.
- Product Code
- HET
- PMA / PMN Number
- K100280
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- OTHER
Narratives
(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE BLADE WAS NOT SEIZED AND THE DEVICE OPERATED AS INTENDED DURING THE EVALUATION.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC MYOMECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE BLADE GOT STUCK AND WOULD NOT ROTATE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352236 | GYNECARE MORCELLEX TISSUE MORCELLATOR | LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) | HET | ETHICON INC. | UNK | MT217335 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |