FDA Adverse Event Malfunction Summary report: N

CONTOUR CURVED CUTTER STAPLER

MDR report key: 3252183 · Received July 29, 2013

Report

Report Number
3005075853-2013-03785
Event Type
Malfunction
Date Received
July 29, 2013
Date of Event
July 2, 2013
Report Date
July 2, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K040038
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THERE WAS AN INCOMPLETE STAPLE LINE. NO STAPLES CAME OUT IN THE PROXIMAL AND DISTAL PART, ONLY MIDDLE OF THE STAPLE LINE WAS COMPLETE. THE SURGEON USED HAND SEWING TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352590 CONTOUR CURVED CUTTER STAPLER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK K4C04X

Patients

Seq Age Sex Outcome Treatment
1