FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 3252011 · Received July 29, 2013

Report

Report Number
2531779-2013-12010
Event Type
Malfunction
Date Received
July 29, 2013
Report Date
July 8, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/20/2013 WITH THE FOLLOWING FINDINGS: NO DAMAGE WAS OBSERVED TO THE PUMP KEYPAD COVER. DURING TESTING, ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE PRESSES TO ENGAGE. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2013 AND REPORTED THE OK KEYPAD BUTTON WAS UNRESPONSIVE. THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD RESPONSE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352445 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1