FDA Adverse Event
Summary report: N
BRILLIANCE BIGBORE ONCOLOGY
MDR report key: 3251837
·
Received July 25, 2013
Report
- Report Number
- 1525965-2013-00204
- Date Received
- July 25, 2013
- Report Date
- June 27, 2013
- Manufacturer
- PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
- Product Code
- JAK
- PMA / PMN Number
- K033357
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NOTE: WE HAVE NOT COMPLETED OUR INVESTIGATION OF THIS EVENT. WE WILL FILE A F/U MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4).
Description of Event or Problem · 1
PHILIPS RECEIVED INFO WHERE THE CUSTOMER REPORTED THAT THE COUCH WILL NOT ZERO AND THERE IS NO TILTING POSSIBLE WITH THE GANTRY. THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE SYSTEM AND FOUND A REPORT OF A "STUCK BUTTON" IN THE ERROR LOG FILES. THE FSE DETERMINED THAT THE MULTI-FUNCTION FOOTSWITCH HAD MALFUNCTIONED WHICH CAUSED THE ERROR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 347838 | BRILLIANCE BIGBORE ONCOLOGY | JAK | PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC. | 728243 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |