ACCU-CHEK SPIRIT COMBO
Report
- Report Number
- 2183996-2013-01412
- Event Type
- Malfunction
- Date Received
- July 29, 2013
- Date of Event
- July 24, 2013
- Report Date
- October 29, 2013
- Manufacturer
- ROCHE HEALTH SOLUTIONS, INC
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
CONCLUSION - THE COMPLAINT CAN BE VERIFIED. RESULT THE MENU BUTTON DOES NOT RESPOND. THERE ARE PARTICLES OF PLASTIC INSIDE THE HOUSING OF THE MENU BUTTON. THE PARTICLES TEMPORARY ISOLATE THE AREA OF CONTACT INSIDE THE BUTTON HOUSING. DUE TO THIS PROBLEM THE MENU BUTTON DOES NOT RESPOND AND IS WITHOUT FUNCTION.
THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
PATIENT REPORTED THE MENU BUTTON ON THE INFUSION DEVICE DOES NOT WORK CORRECTLY. PATIENT REPORTED HAVING TO PRESS THE BUTTON REPEATEDLY AND THE BUTTON DOES NOT ALWAYS RESPOND. NO FURTHER INFORMATION IS AVAILABLE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. REQUESTED RETURN OF THE ALLEGED INFUSION DEVICE FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352886 | ACCU-CHEK SPIRIT COMBO | INSULIN INFUSION PUMP | LZG | ROCHE HEALTH SOLUTIONS, INC | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |