FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3251520 · Received July 29, 2013

Report

Report Number
2531779-2013-11995
Event Type
Malfunction
Date Received
July 29, 2013
Report Date
July 3, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING EVALUATION, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN WAS FOUND TO BE DIM. A TEST SCREEN WAS INSTALLED AND DISPLAYED NORMALLY. UNRELATED TO THE DISPLAY ISSUE, THE KEYPAD COVER WAS COMPLETELY PEELED OFF AND THE CONTRAST BUTTON WAS MISSING. THE CONTRAST BUTTON WAS INOPERABLE; ALL OTHER BUTTONS RESPONDED APPROPRIATELY. EVIDENCE OF CONTAMINATION WAS FOUND UNDER THE UP ARROW, DOWN ARROW, AND OK KEY CONTACTS.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DIM DISPLAY SCREEN. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352495 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR