PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3251214
·
Received July 29, 2013
Report
- Report Number
- 2032227-2013-03178
- Event Type
- Injury
- Date Received
- July 29, 2013
- Date of Event
- July 1, 2013
- Report Date
- July 2, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT HE IS GOING TO THE HOSPITAL DUE TO HIGH BLOOD GLUCOSE. CUSTOMER BLOOD GLUCOSE READING WAS OVER 300 MG/DL. CUSTOMER STATED THAT THE INFUSION SET CANNULA WAS BENT AND THAT THE INSULIN PUMP WAS ALARMING NO DELIVERY. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352538 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR | Hospitalization |