UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM
Report
- Report Number
- 2122870-2013-00644
- Event Type
- Malfunction
- Date Received
- July 29, 2013
- Date of Event
- July 4, 2013
- Report Date
- July 7, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- MMI
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) REPLACED THE SUBSTRATE VALVE AND NOTICED LOOSE SUBSTRATE CAP FITTINGS. THE FSE TIGHTENED THE FITTINGS. THE FSE PERFORMED SYSTEM CHECK, 50 REPETITIONS OF LUMIBLANK TO TEST SUBSTRATE DELIVERY, CALIBRATED TROPONIN ASSAY, AND COMPLETED ALL LEVELS OF QUALITY CONTROL (QC). SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. IN CONCLUSION, A MALFUNCTION OF THE SUBSTRATE DELIVERY SYSTEM IS THE LIKELY CAUSE OF THIS EVENT. IMPROPER DISPENSING OF SUBSTRATE INTO THE REACTION VESSELS (RVS) MAY LEAD TO DECREASED RELATIVE LIGHT UNITS (RLU'S) VALUES AND THE NO VALUE/IND FLAGGED PATIENT RESULTS. THIS MEDWATCH REPORT IS RELATED TO MDRS: 2122870-2013-00645 AND 2122870-2013-00646
THE CUSTOMER REPORTED NO VALUE AND IND (INDETERMINATE) FLAGGED TROPONIN I (ACCESS ACCUTNI) RESULTS, FOR MULTIPLE PATIENTS, INVOLVING THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM USED IN CONJUNCTION WITH THE ACCESS ACCUTNI ASSAY AND CALIBRATOR. THE CUSTOMER STATED THE NON-NUMERICAL RESULTS WERE NOT REPORTED FROM THE LABORATORY AS THESE WERE NOT ACTUAL PATIENT VALUES. THERE WAS NO PATIENT INJURY OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER STATED THE PATIENTS' SAMPLES WERE RETESTED ON THE LABORATORY'S ALTERNATE ACCESS 2 ANALYZER; ACTUAL RESULTS WERE NOT SUPPLIED. THE CUSTOMER PERFORMED A REAGENT LOT-TO-LOT COMPARISON OF PATIENT SAMPLES FOR HUMAN CHORIONIC GONADOTROPIN (HCG) AND THE RESULTS WERE NOT REPRODUCIBLE AND OUTSIDE THE ACCEPTABLE LIMITS. ALL LEVELS OF QUALITY CONTROL (QC) RECOVERED WITHIN THE LABORATORY'S ESTABLISHED RANGES. CALIBRATION PASSED WITH ACCEPTABLE PERCENT COEFFICIENT OF VARIATION (%CV). SERVICE WAS REQUESTED AND A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THIS IS REPORT ONE OF THREE REFERENCING THE PATIENT ON THE EVENT DATE NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 353583 | UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | MMI | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |