FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3251091 · Received July 29, 2013

Report

Report Number
3004209178-2013-96152
Event Type
Malfunction
Date Received
July 29, 2013
Date of Event
July 3, 2013
Report Date
July 3, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. THE DISPLAY FUNCTIONS PROPERLY AND NO RAINBOW EFFECT ON THE DISPLAY WAS NOTED. THE DEVICE HAD MISSING SEGMENTS ON DISPLAY OR OVERHEATS DISPLAY NOTED. THE UNIT HAD SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP DISPLAY DID NOT RETURN TO NORMAL AFTER CHANGING THE BATTERY. THE CALLER ALSO STATED THAT THE INSULIN PUMP WAS DELIVERING MORE INSULIN. THE BLOOD GLUCOSE READING AT TIME OF CALL WAS 180MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352290 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 23 YR