FDA Adverse Event Malfunction Summary report: N

COULTER® LH 780 HEMATOLOGY ANALYZER

MDR report key: 3251074 · Received July 29, 2013

Report

Report Number
1061932-2013-01564
Event Type
Malfunction
Date Received
July 29, 2013
Date of Event
July 9, 2013
Report Date
July 9, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE. THE FSE DISCOVERED A WORN LOWER SHEATH RESTRICTOR TUBING AT THE LOCATION OF CONTACT WITH THE DIFFERENTIAL MIXING ASSEMBLY. THE FSE REPLACED THE RESTRICTOR TUBING TO RESOLVE THE LEAK AND VERIFIED REPAIR PER ESTABLISHED PROCEDURES. FAILURE MODE OF THE LEAK IS ATTRIBUTED TO A WORN SHEATH RESTRICTOR TUBING. BECKMAN COULTER INTERNAL IDENTIFIER FOR THIS REPORT IS (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A LEAK FROM A DIFFERENTIAL MIXING CHAMBER WHICH APPEARED CRACKED ON THE COULTER LH 780 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS UNDETERMINED BUT THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING GLOVES AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO ERRONEOUS PATIENT RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353664 COULTER® LH 780 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA