FDA Adverse Event Malfunction Summary report: N

FOLFUSOR

MDR report key: 3250897 · Received July 29, 2013

Report

Report Number
1416980-2013-19947
Event Type
Malfunction
Date Received
July 29, 2013
Report Date
July 3, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. AS THE SAMPLE WAS NOT RETURNED, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ELASTOMERIC BALLOON OF A FOLFUSOR LV5 DEVICE DID NOT FILL WITH SOLUTION AND INSTEAD THE SOLUTION FLOWED INTO THE HOUSING. THIS OCCURRED WHILE FILLING THE DEVICE WITH A SALINE SOLUTION USING A LUER-LOCK SYRINGE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352648 FOLFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12N016

Patients

Seq Age Sex Outcome Treatment
1