FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3250829
·
Received July 29, 2013
Report
- Report Number
- 1416980-2013-19936
- Event Type
- Malfunction
- Date Received
- July 29, 2013
- Date of Event
- July 1, 2013
- Report Date
- July 3, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PM
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). (B)(6). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION OF THE DEVICE IDENTIFIED THAT THE DOOR LATCH WAS BROKEN, CONFIRMING THE CUSTOMER REPORTED CONDITION. TO CORRECT THE CONDITION, THE DOOR LATCH ASSEMBLY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP HAD A DAMAGED DOOR. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 353137 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |