SYSTEM 6 ASEPTIC HOUSING ASSY
Report
- Report Number
- 0001811755-2013-01773
- Event Type
- Malfunction
- Date Received
- July 26, 2013
- Date of Event
- June 28, 2013
- Report Date
- July 1, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE DEVICE HAS BEEN RECEIVED, BUT THE EVALUATION HAS NOT YET BEGUN. ADDITIONAL INFORMATION MAY BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETE. FAILURE ANALYSIS IS ONGOING; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE RESULTS ANALYSIS IS COMPLETE.
THE REPORTED EVENT, OPENED DURING SURGERY, WAS DUPLICATED; IT WAS CONFIRMED THE DELRIN BALL DETENT WAS MISSING DURING VISUAL INSPECTION OF THE DEVICE. A MISSING DELRIN BALL MAY ALLOW THE LID OF THE HOUSING TO OPEN UNINTENTIONALLY, WHICH IS A PROBABLE CAUSE OF THE REPORTED EVENT. THE DEVICE WAS DISCARDED BY THE MANUFACTURER.
IT WAS REPORTED THAT THE DELRIN BALL IN THE LOCKING MECHANISM OF THE ASEPTIC BATTERY HOUSING WAS MISSING OR FALLING OUT. NOTHING FELL INTO A SURGICAL SITE, AS THIS WAS DETERMINED WHILE INSPECTING THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
IT WAS REPORTED THAT THE DELRIN BALL IN THE LOCKING MECHANISM OF THE ASEPTIC BATTERY HOUSING WAS MISSING OR FALLING OUT. NOTHING FELL INTO A SURGICAL SITE, AS THIS WAS DETERMINED WHILE INSPECTING THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350034 | SYSTEM 6 ASEPTIC HOUSING ASSY | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | KIJ | STRYKER INSTRUMENTS-KALAMAZOO | 09078 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |