FDA Adverse Event Malfunction Summary report: N

PROXIMATE ILS CURVED CIR STAPL

MDR report key: 3250516 · Received July 26, 2013

Report

Report Number
3005075853-2013-03769
Event Type
Malfunction
Date Received
July 26, 2013
Date of Event
June 28, 2013
Report Date
July 3, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ATTEMPTS HAVE BEEN MADE TO HAVE DEVICE RETURNED. DEVICE LOCATION IS UNKNOWN.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT BEFORE FIRING THE DEVICE THE ORANGE INDICATOR SHOULD BE FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE. IN ADDITION IT SHOULD BE NOTED THAT IF THE FIRING SEQUENCE IS NOT COMPLETE (THE FIRING HANDLE REACHES ITS STOPPING POINT, AND THE FIRING TRIGGER IS PARALLEL TO THE INSTRUMENT HANDLE) STAPLES COULD BE PARTIALLY DEPLOYED WITHOUT CUTTING THE WASHER. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE COULD NOT BE FIRED. THE FIRING HANDLE COULD NOT BE GRASPED FULLY AND THE WASHER COULD NOT BE PUNCHED OUT. A NEW DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350940 PROXIMATE ILS CURVED CIR STAPL STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1