PROXIMATE ILS CURVED CIR STAPL
Report
- Report Number
- 3005075853-2013-03769
- Event Type
- Malfunction
- Date Received
- July 26, 2013
- Date of Event
- June 28, 2013
- Report Date
- July 3, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). ATTEMPTS HAVE BEEN MADE TO HAVE DEVICE RETURNED. DEVICE LOCATION IS UNKNOWN.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT BEFORE FIRING THE DEVICE THE ORANGE INDICATOR SHOULD BE FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE. IN ADDITION IT SHOULD BE NOTED THAT IF THE FIRING SEQUENCE IS NOT COMPLETE (THE FIRING HANDLE REACHES ITS STOPPING POINT, AND THE FIRING TRIGGER IS PARALLEL TO THE INSTRUMENT HANDLE) STAPLES COULD BE PARTIALLY DEPLOYED WITHOUT CUTTING THE WASHER. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4).
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE COULD NOT BE FIRED. THE FIRING HANDLE COULD NOT BE GRASPED FULLY AND THE WASHER COULD NOT BE PUNCHED OUT. A NEW DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350940 | PROXIMATE ILS CURVED CIR STAPL | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |