FDA Adverse Event Other Summary report: N

CK COR-KNOT

MDR report key: 3250473 · Received May 22, 2013

Report

Report Number
1320468-2013-00002
Event Type
Other
Date Received
May 22, 2013
Date of Event
March 29, 2013
Report Date
April 3, 2013
Manufacturer
LSI SOLUTIONS INC.
Product Code
GAW
PMA / PMN Number
K100593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTING FACILITY DID NOT IDENTIFY THE REPORTED CK COR-KNOT DEVICE. THE REPORTING FACILITY RETURNED FOUR DEVICES. THE MANUFACTURER'S ENGINEER EXAMINED THE FOUR DEVICES AND THE REPORT THAT THE "DEVICE FOR CINCHING VALVE SUTURE MISFIRED WHILE BEING DEPLOYED" COULD NOT BE REPLICATED. THE DEVICES WERE VISUALLY AND FUNCTIONALLY EXAMINED AND THE DEVICES MET THE SPECIFICATION REQUIREMENTS. A REVIEW OF TWO YEARS OF COMPLAINTS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS. THE FAILURE COULD NOT BE CONFIRMED ON THE REPORTED DEVICE. THEREFORE, NO REASONABLE OR DEFINED COURSE OF ACTION IS AVAILABLE BASED ON LACK OF EVIDENCE OF A CONFIRMED FAILURE MODE.

Description of Event or Problem · 1

PLEASE REFER IMPORTER REPORT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
226569 CK COR-KNOT SUTURE, NONABSORBABLE, SYNTHETIC GAW LSI SOLUTIONS INC. 14955

Patients

Seq Age Sex Outcome Treatment
1 80 YR Other