FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 3250366 · Received July 26, 2013

Report

Report Number
0001811755-2013-01768
Event Type
Malfunction
Date Received
July 26, 2013
Date of Event
June 18, 2013
Report Date
June 18, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
FYD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT WAS CONFIRMED; THIS WAS DETERMINED TO BE DUE TO THE BROKEN USER INTERFACE TOWER AND USER INTERFACE PANEL. THE TECHNICIAN REPLACED THE INTERFACE PANEL AND RETURNED THE UNIT TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, EXPOSED WIRES WERE FOUND AT THE INTERFACE PANEL OF THE NEPTUNE GOLD ROVER. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, EXPOSED WIRES WERE FOUND AT THE INTERFACE PANEL OF THE NEPTUNE GOLD ROVER. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349999 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE APPARATUS, EXHAUST, SURGICAL FYD STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1