FDA Adverse Event Malfunction Summary report: N

FUSION OMNI-TOME

MDR report key: 3250118 · Received July 23, 2013

Report

Report Number
1037905-2013-00570
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
May 16, 2013
Report Date
June 24, 2013
Manufacturer
WILSON-COOK MEDICAL INC.
Product Code
KNS
PMA / PMN Number
K052051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THAT THE CUTTING WIRE HAS DISCONNECTED FROM THE CATHETER AT THE DISTAL END. THE CUTTING WIRE IS INTACT AND REMAINS SECURELY ATTACHED TO THE SPHINCTEROTOME AT THE PROXIMAL END. HOWEVER, DUE TO THE CATHETER AND SECURING COMPONENT DISCONNECTION, THE DISTAL END OF THE CUTTING WIRE IS NO LONGER CONNECTED TO THE SPHINCTEROTOME CATHETER AT THE DISTAL END. NO SECTION OF THE SPHINCTEROTOME DEVICE IS MISSING. THE CUTTING WIRE EXHIBITS EVIDENCE OF A CAUTERY APPLICATION (BLACKENING OF THE CUTTING WIRE WAS NOTED). A FURTHER EXAMINATION OF THE CATHETER TIP WAS CONDUCTED AND THE OPENING OF THE CATHETER AT HE LOCATION OF THE DISCONNECTION IS EXTENSIVELY TORN. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER CAN OCCUR IF THE TIP OF THE SPHINCTEROTOME IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER NOT TO OVER FLEX OR BOW THE TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE THE SPHINCTEROTOME. THIS TYPE OF DAMAGE CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRE-CURVED AND IS PROVIDED WITH A PRE-CURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF THE SPHINCTEROTOME TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE." IF THE ELEVATOR OF THE ENDOSCOPE REMAINS IN THE CLOSED/ UP POSITION WHEN RETRACTION OF THE SPHINCTEROTOME IS ATTEMPTED AND ADDITIONAL PRESSURE IS APPLIED, THIS COULD HAVE CONTRIBUTED TO SEPARATION OF THE CATHETER AND CUTTING WIRE SECURING COMPONENT. THE INSTRUCTIONS FOR USE CAUTION THE USER THE ELEVATOR SHOULD REMAIN OPEN/ DOWN WHEN ADVANCING OR RETRACTING THE SPHINCTEROTOME. THIS ACTIVITY WILL AID IN DEVICE PRESERVATION. OTHER FACTORS THAT CAN CONTRIBUTE TO SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME UPON REMOVING THE DEVICE FROM PACKAGING. THE USER IS THEN INSTRUCTED TO CAREFULLY REMOVE THE PRECURVED STYLET FROM THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. THE FUNCTIONAL TEST INCLUDES BOWING THE SPHINCTEROTOME TO ENSURE THE DISTAL END RESPONDS TO HANDLE MANIPULATION. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

ON (B)(6) 2013 THE FOLLOWING INFORMATION WAS PROVIDED. DURING AN ENDOSCOPY PROCEDURE, A COOK FUSION OMNI-TOME WAS USED. THE CUTTING WIRE WAS BROKEN, NO SECTION OF THE DEVICE DETACHED INSIDE THE PATIENT. THIS DID NOT REASONABLY SUGGEST A REPORTABLE EVENT HAD OCCURRED. ON (B)(6) 2013 THE PRODUCT WAS RECEIVED AT COOK. OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE CUTTING WIRE SECURING COMPONENT LOCATED NEAR THE DISTAL END OF THE SPHINCTEROTOME HAS DISCONNECTED FROM THE CATHETER. THIS HAS BEEN ESTABLISH AS A REPORTABLE OCCURRENCE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344400 FUSION OMNI-TOME KNS WILSON-COOK MEDICAL INC. W3248861

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS TJF ENDOSCOPE