FDA Adverse Event
Malfunction
Summary report: N
HUT EXT DR FINAL ASSY-STANDARD
MDR report key: 3250053
·
Received July 23, 2013
Report
- Report Number
- 1518293-2013-00175
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- July 9, 2013
- Report Date
- July 23, 2013
- Manufacturer
- LIEBEL-FLARSHEIM CO
- Product Code
- IXR
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FIELD SVC ENGINEER (FSE) TROUBLESHOT LOSS OF FLUORO AND FOUND A BROKEN WIRING HARNESS. FSE REPLACED THE WIRING HARNESS AND VERIFIED PROPER OPERATION PER (B)(4). SYS RETURNED TO FULL SVC BY THE CUSTOMER.
Description of Event or Problem · 1
ON (B)(6) 2013: CUSTOMER STATES VIA PHONE THAT ALL TABLE FUNCTIONS FAILED AND THE TUBE CARRIAGE ARM WAS STUCK DURING A CYSTO PROCEDURE ON A MALE PT OF UNK AGE. THE PHYSICIAN MOVED THE PT TO ANOTHER TABLE AND COMPLETED THE PROCEDURE WITH A PORTABLE C-ARM FLUORO. NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343253 | HUT EXT DR FINAL ASSY-STANDARD | IXR | LIEBEL-FLARSHEIM CO | HUT EXT DR | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |