FDA Adverse Event Malfunction Summary report: N

HUT EXT DR FINAL ASSY-STANDARD

MDR report key: 3250053 · Received July 23, 2013

Report

Report Number
1518293-2013-00175
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
July 9, 2013
Report Date
July 23, 2013
Manufacturer
LIEBEL-FLARSHEIM CO
Product Code
IXR
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FIELD SVC ENGINEER (FSE) TROUBLESHOT LOSS OF FLUORO AND FOUND A BROKEN WIRING HARNESS. FSE REPLACED THE WIRING HARNESS AND VERIFIED PROPER OPERATION PER (B)(4). SYS RETURNED TO FULL SVC BY THE CUSTOMER.

Description of Event or Problem · 1

ON (B)(6) 2013: CUSTOMER STATES VIA PHONE THAT ALL TABLE FUNCTIONS FAILED AND THE TUBE CARRIAGE ARM WAS STUCK DURING A CYSTO PROCEDURE ON A MALE PT OF UNK AGE. THE PHYSICIAN MOVED THE PT TO ANOTHER TABLE AND COMPLETED THE PROCEDURE WITH A PORTABLE C-ARM FLUORO. NO PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343253 HUT EXT DR FINAL ASSY-STANDARD IXR LIEBEL-FLARSHEIM CO HUT EXT DR NA

Patients

Seq Age Sex Outcome Treatment
1 UNK