COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER
Report
- Report Number
- 1061932-2013-01500
- Event Type
- Malfunction
- Date Received
- July 26, 2013
- Date of Event
- June 28, 2013
- Report Date
- June 28, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K010765
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED AND OBSERVED A LEAK FROM A DISCONNECTED TUBING AT THE FOAM TRAP FMT3. THE FSE REPLACED THE AFFECTED TUBING TO RESOLVE THE LEAK AND COMPLETED THE PM (PREVENTIVE MAINTENANCE). THE FSE STARTED UP THE INSTRUMENT TO VERIFY REPAIRS AND OBSERVED THAT THE WIRES TO THE WBC (WHITE BLOOD CELL) OVERFLOW SENSOR WERE BROKEN. THE FSE REPLACED THE WBC OVERFLOW SENSOR AND NO FURTHER LEAKS WERE OBSERVED. FAILURE MODE OF THE LEAK AND VACUUM ERRORS IS ATTRIBUTED TO A DISCONNECTED RESTRICTOR TUBING AT THE FOAM TRAP FMT3. THE FSE ALSO DISCOVERED THAT THE WBC OVERFLOW SENSOR WERE BROKEN. BECKMAN COULTER INTERNAL IDENTIFIER FOR THIS REPORT IS (B)(4).
THE CUSTOMER REPORTED OBSERVING A LEAK COMING FROM THE FOAM TRAP OF THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER WHILE TROUBLESHOOTING FOR "LOW VACUUM OUT OF RANGE" ERRORS. THE CUSTOMER INDICATED THAT APPROXIMATELY 1 - 2 ML OF FLUID LEAKED BUT WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES, A LABORATORY COAT, AND EYEGLASSES AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO ERRONEOUS PATIENT RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 351081 | COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |